Máy Phân Tích Khí

2700
COMBUSTION GAS ANALYSER

For continuous measurement of oxygen and/or combustibles (COe) and enable increased combustion efficiency

• Unique close coupled, low extracted volume design protects sensors and provides fast, in-phase O2 and COe measurement without probe blockages
• Long life, low drift zirconia and catalytic sensors
• Protective, BASEEFA tested flametraps as standard
• Wide variety of probes and flanges
• Configurations for safe or hazardous area use
• Proven applications include power generation boilers, process heaters, thermal crackers and incinerators

 
SPECIFICATION
 
GASES MEASURED
Oxygen (O2)
Combustibles (COe - Carbon Monoxide Equivalent)
TECHNOLOGY
Zirconium oxid (zirconia)
Patented thick film catalytic sensor
Typical fuels
All
<2.5% sulphur gas, oil and coal 1
PERFORMANCE f
 
 
Min. recommended range:
Max. COe in stream:
0-1% O2
Not applicable 
0-500ppm
6000ppm  
Intrinsic error (accuracy):
Recommended resolution:
Sensor head response time (T90):
Typical calibration period:
±1% of reading or ±0.1% O2*
0.01% O2
<10 seconds
12 months
±25ppm or ±5% of reading*
1ppm
<20 seconds
2 months
  
SIGNAL INPUTS/OUTPUTS
 
 
Analogue Outputs:                   One configurable isolated 0/4-20mA per measurement (impedance 600
                                                                 Ω or less)
O2 output configurable from 0-1% min. to 0-25% max. O2 in 1% steps
COe output configurable from 0-500ppm to 0-15,000ppm (measurement range remains  as above)
Alarms & Relays:                    Four SPCO relays (250V ac/3A or 28V dc/1A max), configurable for Concentration Alarms, Faults, In Calibration, In Blowback, and solenoid valves for blowback and autocalibration
Digital Inputs:                          Two non-isolated digital inputs provided to initiate autocalibration and
                                                             perform blowback
PHYSICAL
Ingress Protection:                   IP66/NEMA 4X
Weight:                                   Control Unit: <11kg (<24.3lbs) Sensor Head, dual transducer: <17kg
                                                          (<37.5lbs)
Mounting:                                Wall mounting sensor head, and wall mount 19” rack or panel mount
                                               control unit, with a wide choice of probes,mounting flanges and
                                                              adaptors to assist retrofitting. See over for dimensions and more details
ENVIRONMENTAL
   Area classification:                Versions available for safe and hazardous  areas (see over for details)
Ambient temperature:             Sensor head -20°C to +70°C (-4°F to +158°F)
                                             Control unit -10°C to +55°C (+14°F to +131°F)
  Altitude:                                 Maximum 2000m (6500ft)
 
clip_image001[4]clip_image002clip_image003clip_image004clip_image005clip_image006* whichever is greater
1 Consult Servomex for high sulphur (> 2.5%) and other applications
***********************************************************************************************************************
Blog: Máy phân tích khí -----||----- Gas Analyzer
***********************************************************************************************************************

0 nhận xét:

Đăng nhận xét

 
Máy Phân Tích Khí © 2013. All Rights Reserved. Powered by Máy Phân Tích Khí
Top