The Servomex 1800 analyser is a stable, accurate and highly specific oxygen analyser for safe areas.
Designed for safe area oxygen analysis
Low maintenance and re-calibration requirements
Range of alarm outputs to aid integration with other systems
Easy to set up and operate
Special versions for solvent bearing samples
Special high flow rate cell option
Specification |
|
Gases Measured: | O2 |
PERFORMANCE Technology: Range: Intrinsic Error (accuracy): Linearity: Repeatability: Response time (T90) Zero drift/ week: Span drift/ week: SIGNAL OUTPUTS Analogue: Alarms: Concentration Sample flow fail
PHYSICAL Dimensions:(W x D x H) Weight: Hazardous area Classification: Ingress Protection: Mounting: |
Paramagnetic transducer 0-100% O2 <0.2% of reading or 0.05% O2 <0.05% O2 <0.1% of reading or 0.05% O2 4 to 8 seconds <0.05% O2 <1% of reading or 0.05% O2
One 4-20 mA, isolated & one 0-1V dc, non-isolated. Ranges selectable from 0 - 2.5, 5, 10, 25 and 100% O2 Two volt free changeover relays rated 250Vac/3A or 28Vdc (non inductive) maximum and 5V/10mA ac/dc minimum One volt free changeover relay rated 250Vac/3A or 28Vdc (non inductive) maximum and 5V/10mA ac/dc minimum 448mm x 229mm x 235 mm / 17.6” x 9” x 9.25” 26 kg / 57 lbs Non-Hazardous Areas only IP 66 / NEMA 4X Wall or Panel |